2-8
Automated Verification
Step 3. Device Verification
of the verification mismatch to reference test port 1.
Figure 2-6. Mismatch Device Verification Setup
7
Press
RETURN
,
TESTS
,
30,
x1
,
EXECUTE TEST
,
CONTINUE
.
8
When the Agilent 8702D finishes all the measurements, connect the mismatch
verification device, as shown in
. Notice that Port B is now connected
to reference test port 1.
Figure 2-7. Mismatch Device Verification Setup
9
Press
RETURN
,
TESTS
,
31
,
x1
,
EXECUTE TEST
,
CONTINUE
.