![Agilent Technologies 8702D Installation Manual Download Page 164](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953164.webp)
5-20
Automated Verification – Option 011
In Case of Difficulty
8
When the analyzer finishes all the measurements, reconnect the mismatch
verification device, as shown in
.
Figure 5-14. Mismatch Device Verification Setup
9
Press
RETURN
,
TESTS
,
31
,
x1
,
EXECUTE TEST
,
CONTINUE
.
10
You have completed the system verification procedure when the analyzer
displays,
Ver Def 4 PASS
.
In Case of Difficulty
1
Inspect all connections. Do
not
disconnect the cables from the analyzer test
ports. Doing so will invalidate the calibration that you have done earlier.
2
Recall the measurement calibration and repeat the procedure:
a
Press
PRESET
,
SAVE/RECALL
,
SELECT DISK
,
INTERNAL DISK
,
RETURN
.
b
Using the front panel knob, highlight the title of the one-path 2-port calibra-
tion that you have done earlier.
Summary of Contents for 8702D
Page 1: ...Agilent 8702D Lightwave Component Analyzer Installation Guide ...
Page 10: ......
Page 46: ......
Page 89: ...4 33 Performance Tests 8 Test Port Crosstalk Figure 4 19 Agilent 8702D Bottom View ...
Page 166: ......