![Agilent Technologies 8702D Installation Manual Download Page 237](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953237.webp)
6-71
Performance Tests – Option 011
Performance Test Records - Option 011
+ 1
+ 9
________
________
________
± 0.25
± 0.15
+ 3
+ 7
________
________
________
± 0.25
± 0.15
+ 5
+ 5
________
________
________
± 0.25
± 0.15
+ 7
+ 3
________
________
________
± 0.25
± 0.15
+ 9
+ 1
________
________
________
± 0.25
± 0.15
+ 11
– 1
________
________
________
± 0.25
± 0.15
+ 13
– 3
________
________
________
± 0.25
± 0.15
+ 15
– 5
________
________
________
± 0.5
± 0.15
+ 17
– 7
________
________
________
± 0.5
± 0.15
+ 20
– 10
________
________
________
± 0.5
± 0.15
CW Frequency = 3 GHz
– 5
+ 15
________
________
________
± 0.25
± 0.15
– 3
+ 13
________
________
________
± 0.25
± 0.15
– 1
+ 11
________
________
________
± 0.25
± 0.15
+ 1
+ 9
________
________
________
± 0.25
± 0.15
+ 3
+ 7
________
________
________
± 0.25
± 0.15
+ 5
+ 5
________
________
________
± 0.25
± 0.15
+ 7
+ 3
________
________
________
± 0.25
± 0.15
+ 9
+ 1
________
________
________
± 0.25
± 0.15
+ 11
– 1
________
________
________
± 0.25
± 0.15
+ 13
– 3
________
________
________
± 0.25
± 0.15
+ 15
– 5
________
________
________
± 0.5
± 0.15
+ 17
– 7
________
________
________
± 0.5
± 0.15
Test 2. Source Power Range, Linearity, and Accuracy (Power Range and Power Linearity) (2 of 3)
Source Power Level
(dBm)
Power
Offset
(dB)
Path Loss
(dB)
Measured
Value (dB)
Power
Linearity
(dB)
Specification
(dB)
Meas
Uncertainty