![Agilent Technologies 8702D Installation Manual Download Page 68](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953068.webp)
4-12
Performance Tests
5. Minimum R Channel Level
8
If the analyzer displays a phase lock error message, continue increasing the test
port power until phase lock is achieved.
9
Write the test port power, that is displayed on the analyzer, on the
"Performance Test Record."
10
through
for the other CW frequencies listed in the
"Performance Test Record."
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
Check the flexible RF cable (W8, as shown in
) between the R
sampler assembly (A4) and the A11 phase lock assembly. Make sure it is
connected between A11J1 (PL IF IN) and 1st IF Out.
Summary of Contents for 8702D
Page 1: ...Agilent 8702D Lightwave Component Analyzer Installation Guide ...
Page 10: ......
Page 46: ......
Page 89: ...4 33 Performance Tests 8 Test Port Crosstalk Figure 4 19 Agilent 8702D Bottom View ...
Page 166: ......
Page 256: ......