5-8
Automated Verification – Option 011
Step 3. Device Verification
18
Press
TRANSMISSION
,
FWD TRANS THRU
.
19
When the analyzer finishes the measurement, press
FWD MATCH THRU
.
20
When the analyzer finishes the measurement, press
REV TRANS THRU
.
21
When the analyzer finishes the measurement, press
REV MATCH THRU
.
22
When the analyzer displays,
PRESS ’DONE’ IF FINISHED WITH STD(s)
,
press
STANDARDS DONE
,
DONE 2-PORT CAL
.
23
Press
SAVE/RECALL
,
SELECT DISK
,
INTERNAL MEMORY
,
RETURN
,
SAVE STATE
to save
the calibration into the analyzer’s internal memory.
24
When the analyzer finishes storing the measurement calibration, press
SELECT
DISK
,
INTERNAL DISK
.
Step 3. Device Verification
1
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
28
,
x1
,
EXECUTE TEST
.
2
At the prompt, connect the 20 dB attenuator (supplied in the verification kit)
as shown in
.
Figure 5-4. Connections for the 20 dB Verification Device