4-28
Performance Tests
7. Test Port Input Frequency Response
Figure 4-16. Setup for Power Meter Calibration on Test Port 1
45
Press
MEAS
,
INPUT PORTS
,
TESTPORT 1
.
46
Press
CAL
,
PWRMTR CAL
,
ONE SWEEP
,
TAKE CAL SWEEP
to start the power meter
calibration for output test port 1.
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz
47
When the analyzer displays the message,
POWER METER CALIBRATION
SWEEP DONE
, connect the equipment as shown as in
.