![Agilent Technologies 2000 X User Manual Download Page 97](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_user-manual_2868572097.webp)
Digital Channels
6
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
97
Increasing the ground inductance (L), increasing the current (di) or
decreasing the transition time (dt), will all result in increasing the voltage
(V). When this voltage exceeds the threshold voltage defined in the
oscilloscope, a false data measurement will occur.
Sharing one probe ground with many probes forces all the current that
flows into each probe to return through the same common ground
inductance of the probe whose ground return is used. The result is
increased current (di) in the above equation, and, depending on the
transition time (dt), the common mode voltage may increase to a level that
causes false data generation.
In addition to the common mode voltage,longer ground returns also
degrade the pulse fidelity of the probe system. Rise time is increased, and
ringing, due to the undamped LC circuit at the input of the probe, is also
increased. Because the digital channels display reconstructed waveforms,
they do not show ringing and perturbations. You will not find ground
problems through examination of the waveform display. In fact, it is likely
Figure 12
Common Mode Input Voltage Model
Z
in
Z
in
i
n
i
2
+i
2
i
1
i
1
+i
n
+i
n
Z
in
Probe 1
Probe
Ground
Probe 2
Probe N
L (GND)
Vn (Common Mode)
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...