Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
11
Acquisition Control
Segmented Memory Re-Arm Time
After each segment fills, the oscilloscope re- arms and is ready to trigger
in about 8 µs.
Remember though, for example: if the horizontal time per division control
is set to 5 µs/div, and the Time Reference is set to
Center
, it will take at
least 50 µs to fill all ten divisions and re- arm. (That is 25 µs to capture
pre- trigger data and 25 µs to capture post- trigger data.)
Saving Data from Segmented Memory
You can save either the currently displayed segment (
Save Segment -
Current
), or all segments (
Save Segment - All
) in the following data formats:
CSV, ASCII XY, and BIN.
Be sure to set the Length control to capture enough points to accurately
represent the captured data. When the oscilloscope is busy saving multiple
segments, progress is displayed in the upper right area of the display.
For more information, see
"To save CSV, ASCII XY, or BIN data files"
on
page 220.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...