![Agilent Technologies 2000 X User Manual Download Page 48](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_user-manual_2868572048.webp)
48
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
2
Horizontal Controls
XY Time Mode
The XY time mode converts the oscilloscope from a volts- versus- time
display to a volts- versus- volts display using two input channels. Channel 1
is the X- axis input, channel 2 is the Y- axis input. You can use various
transducers so the display could show strain versus displacement, flow
versus pressure, volts versus current, or voltage versus frequency.
Example
This exercise shows a common use of the XY display mode by measuring
the phase difference between two signals of the same frequency with the
Lissajous method.
1
Connect a sine wave signal to channel 1, and a sine wave signal of the
same frequency but out of phase to channel 2.
2
Press the
[AutoScale]
key, press the
[Horiz]
key; then, press
Time Mode
and select "XY".
3
Center the signal on the display with the channel 1 and 2 position ( )
knobs. Use the channel 1 and 2 volts/div knobs and the channel 1 and
2
Fine
softkeys to expand the signal for convenient viewing.
The phase difference angle (
θ
) can be calculated using the following
formula (assuming the amplitude is the same on both channels):
Figure 3
XY time mode signals, centered on display
sin
θ
=
A
B
or
C
D
A
B
D
C
Signal must
be centered in
“ X”
Measuring
phase difference
Signals 90 degrees
out of phase
Signals
in phase
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...