Mask Testing
14
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
193
On Error
The On Error setting specifies the action(s) to take when the input waveform
does not conform to the mask. This setting supersedes the Run Until setting.
• Stop — The oscilloscope will stop when the first error is detected (on the
first waveform that does not conform to the mask). This setting
supersedes the Minimum # of Tests and Minimum Time settings.
• Save — The oscilloscope saves the screen image when an error is
detected. In the Save Menu (press [Save/Recall] > Save), select an
image format (*.bmp or *.png), destination (on a USB storage device), and
file name (which can be auto-incrementing). If errors occur too frequently
and the oscilloscope spends all its time saving images, press the [Stop]
key to stop acquisitions.
• Print — The oscilloscope prints the screen image when an error is
detected. This option is only available when a printer is connected as
described in
"To print the oscilloscope's display"
on page 231.
• Measure — Measurements (and measurement statistics if your
oscilloscope supports them) run only on waveforms that contain a mask
violation. Measurements are not affected by passing waveforms. This
mode is not available when the acquisition mode is set to Averaging.
Note that you can choose to Print or Save, but you cannot select both at the
same time. All other actions may be selected at the same time. For example,
you can select both Stop and Measure to cause the oscilloscope to measure
and stop on the first error..
You can also output a signal on the rear panel TRIG OUT BNC connector when
there is a mask test failure. See
"Setting the Rear Panel TRIG OUT Source"
on
page 246.
Source Lock
When you turn on Source Lock using the Source Lock softkey, the mask is
redrawn to match the source whenever you move the waveform. For example,
if you change the horizontal timebase or the vertical gain the mask is redrawn
with the new settings.
When you turn off Source Lock, the mask is not redrawn when horizontal or
vertical settings are changed.
Source
If you change the Source channel, the mask is not erased. It is re-scaled to the
vertical gain and offset settings of the channel to which it is assigned. To
create a new mask for the selected source channel, go back up in the menu
hierarchy; then, press Automask, and press Create Mask.
The Source softkey in the Mask Setup Menu is the same as the Source
softkey in the Automask Menu.
Test All
When enabled, all displayed analog channels are included in the mask test.
When disabled, just the selected source channel is included in the test.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...