Triggers
9
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
125
4
Press the sync polarity softkey to set the Video trigger to either positive
(
) or negative (
) sync polarity.
5
Press the
Settings
softkey.
6
In the Video Trigger Menu, press the
Standard
softkey to set the video
standard.
The oscilloscope supports triggering on the following television (TV) and
video standards.
7
Press the
Auto Setup
softkey to automatically set up the oscilloscope for
the selected
Source
and
Standard
:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
•
Trigger type is set to
Video
.
•
Video trigger mode is set to
All Lines
.
•
Display
Grid
type is set to
IRE
(when
Standard
is
NTSC
) or
mV
(see
"To
select the grid type"
on page 104).
N O T E
Provide Correct Matching
Many video signals are produced from 75
Ω
sources. To provide correct matching to these
sources, a 75
Ω
terminator (such as an Agilent 11094B) should be connected to the
oscilloscope input.
Standard
Type
Sync Pulse
NTSC
Interlaced
Bi-level
PAL
Interlaced
Bi-level
PAL-M
Interlaced
Bi-level
SECAM
Interlaced
Bi-level
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...