Acquisition Control
11
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
149
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed, and there is a
maximum sample rate associated with oscilloscope's analog- to- digital
converter; however, the actual sample rate is determined by the time of
the acquisition (which is set according to the oscilloscope's horizontal
time/div scale).
sample rate = number of samples / time of acquisition
For example, when storing 50 µs of data in 50,000 points of memory, the
actual sample rate is 1 GSa/s.
Likewise, when storing 50 ms of data in 50,000 points of memory, the
actual sample rate is 1 MSa/s.
The actual sample rate is displayed in the right- side information area.
The oscilloscope achieves the actual sample rate by throwing away
(decimating) unneeded samples.
Selecting the Acquisition Mode
When selecting the oscilloscope acquisition mode, keep in mind that
samples are normally decimated at slower time/div settings.
At slower time/div settings, the effective sample rate drops (and the
effective sample period increases) because the acquisition time increases
and the oscilloscope's digitizer is sampling faster than is required to fill
memory.
For example, suppose an oscilloscope's digitizer has a sample period of
1 ns (maximum sample rate of 1 GSa/s) and a 1 M memory depth. At that
rate, memory is filled in 1 ms. If the acquisition time is 100 ms
(10 ms/div), only 1 of every 100 samples is needed to fill memory.
To select the acquisition mode:
1
Press the
[Acquire]
key on the front panel.
2
In the Acquire Menu, press the
Acq Mode
softkey; then, turn the Entry
knob to select the acquisition mode.
The InfiniiVision oscilloscopes have the following acquisition modes:
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...