Trigger Mode/Coupling
10
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
137
Trigger Indicator
The trigger indicator at the top right of the display shows whether triggers
are occurring.
In the
Auto
trigger mode, the trigger indicator can show:
•
Auto?
(flashing) — the trigger condition is not found (after the
pre- trigger buffer has filled), and forced triggers and acquisitions are
occurring.
•
Auto
(not flashing) — the trigger condition is found (or the pre- trigger
buffer is being filled).
In the
Normal
trigger mode, the trigger indicator can show:
•
Trig'd?
(flashing) — the trigger condition is not found (after the
pre- trigger buffer has filled), and no acquisitions are occurring.
•
Trig'd
(not flashing) — trigger condition is found (or pre- trigger buffer is
being filled).
When the oscilloscope is not running, the trigger indicator area shows
Stop
.
When to Use Auto
Trigger Mode
The
Auto
trigger mode is appropriate when:
•
Checking DC signals or signals with unknown levels or activity.
•
When trigger conditions occur often enough that forced triggers are
unnecessary.
When to Use
Normal Trigger
Mode
The
Normal
trigger mode is appropriate when:
•
You only want to acquire specific events specified by the trigger
settings.
•
Triggering on an infrequent signal from a serial bus (for example, I2C,
SPI, CAN, LIN, etc.) or another signal that arrives in bursts. The
Normal
trigger mode lets you stabilize the display by preventing the oscilloscope
from auto- triggering.
•
Making single- shot acquisitions with the
[Single]
key.
Often with single- shot acquisitions, you must initiate some action in the
device under test, and you don't want the oscilloscope to auto- trigger
before that happens. Before initiating the action in the circuit, wait for
the trigger condition indicator
Trig'd?
to flash (this tells you the
pre- trigger buffer is filled).
See Also
•
"Forcing a Trigger"
on page 115
•
"To set the trigger holdoff"
on page 140
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...