Acquisition Control
11
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
145
Overview of Sampling
To understand the oscilloscope's sampling and acquisition modes, it is
helpful to understand sampling theory, aliasing, oscilloscope bandwidth
and sample rate, oscilloscope rise time, oscilloscope bandwidth required,
and how memory depth affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth
(band- limited) signal with maximum frequency f
MAX
, the equally spaced
sampling frequency f
S
must be greater than twice the maximum frequency
f
MAX
, in order to have the signal be uniquely reconstructed without
aliasing.
f
MAX
= f
S
/2 = Nyquist frequency (f
N
) = folding frequency
Aliasing
Aliasing occurs when signals are under- sampled (f
S
< 2f
MAX
). Aliasing is
the signal distortion caused by low frequencies falsely reconstructed from
an insufficient number of sample points.
Figure 21
Aliasing
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...