126
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
9
Triggers
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM
standards.
•
Horizontal delay is set so that trigger is at first horizontal division
from the left.
You can also press
[Analyze]> Features
and then select
Video
to quickly
access the video triggering automatic set up and display options.
8
Press the
Mode
softkey to select the portion of the video signal that you
would like to trigger on.
The Video trigger modes available are:
•
Field1
and
Field2
— Trigger on the rising edge of the first serration
pulse of field 1 or field 2 (interlaced standards only).
•
All Fields
— Trigger on the rising edge of the first pulse in the vertical
sync interval.
•
All Lines
— Trigger on all horizontal sync pulses.
•
Line: Field1
and
Line:Field2
— Trigger on the selected line # in field 1 or
field 2 (interlaced standards only).
•
Line: Alternate
— Alternately trigger on the selected line # in field 1
and field 2 (NTSC, PAL, PAL- M, and SECAM only).
9
If you select a line # mode, press the
Line #
softkey, then rotate the
Entry knob to select the line number on which you want to trigger.
The following table lists the line (or count) numbers per field for each
video standard.
Video Triggering
Examples
The following are exercises to familiarize you with video triggering. These
exercises use the NTSC video standard.
•
"To trigger on a specific line of video"
on page 127
Video standard
Field 1
Field 2
Alt Field
NTSC
1 to 263
1 to 262
1 to 262
PAL
1 to 313
314 to 625
1 to 312
PAL-M
1 to 263
264 to 525
1 to 262
SECAM
1 to 313
314 to 625
1 to 312
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...