Acquisition Control
11
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
155
See Also
•
Chapter 10
, “Trigger Mode/Coupling,” starting on page 135
High Resolution Acquisition Mode
In High Resolution mode, at slower time/div settings when decimation
would normally occur, extra samples are averaged in order to reduce
random noise, produce a smoother trace on the screen, and effectively
increase vertical resolution.
High Resolution mode averages sequential sample points within the same
acquisition. An extra bit of vertical resolution is produced for every factor
of 4 averages. The number of extra bits of vertical resolution is dependent
on the oscilloscope's time per division setting (sweep speed) and the
displayed sample rate.
The slower the time/div setting, the greater the number of samples that
are averaged together for each display point.
High Resolution mode can be used on both single- shot and repetitive
signals and it does not slow waveform update because the computation is
done in the MegaZoom custom ASIC. High Resolution mode limits the
oscilloscope's real- time bandwidth because it effectively acts like a
low- pass filter.
Acquiring to Segmented Memory
You can purchase the oscilloscope with the segmented memory option
factory- installed (Option SGM) or you can install a license to enable it
(order model number DSOX2SGM "Segmented Memory").
Displayed Sample Rate (sr, per
channel, 1 Gsa/s max)
Displayed Sample Rate (sr,
interleaved, 2 Gsa/s max)
Bits of
resolution
250 MSa/s < sr
≤
1 Gsa/s
500 MSa/s < sr
≤
2 Gsa/s
8
62.5 MSa/s < sr
≤
250 MSa/s
125 MSa/s < sr
≤
500 MSa/s
9
12.5 MSa/s < sr
≤
62.5 MSa/s
25 MSa/s < sr
≤
125 MSa/s
10
2.5 MSa/s < sr
≤
12.5 MSa/s
5 MSa/s < sr
≤
25 MSa/s
11
sr
≤
2.5 MSa/s
sr
≤
5 MSa/s
12
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...