![Agilent Technologies 2000 X User Manual Download Page 127](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_user-manual_2868572127.webp)
Triggers
9
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
127
•
"To trigger on all sync pulses"
on page 128
•
"To trigger on a specific field of the video signal"
on page 129
•
"To trigger on all fields of the video signal"
on page 130
•
"To trigger on odd or even fields"
on page 131
To trigger on a specific line of video
Video triggering requires greater than 1/2 division of sync amplitude with
any analog channel as the trigger source. Turning the trigger
Level
knob in
Video trigger does not change the trigger level because the trigger level is
automatically set to the sync pulse tips.
One example of triggering on a specific line of video is looking at the
vertical interval test signals (VITS), which are typically in line 18. Another
example is closed captioning, which is typically in line 21.
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select
Video
.
3
Press the
Settings
softkey, then press the
Standard
softkey to select the
appropriate TV standard (NTSC).
4
Press the
Mode
softkey and select the TV field of the line you want to
trigger on. You can choose
Line:Field1
,
Line:Field2
, or
Line:Alternate
.
5
Press the
Line #
softkey and select the number of the line you want to
examine.
N O T E
Alternate Triggering
If Line:Alternate is selected, the oscilloscope will alternately trigger on the selected line
number in Field 1 and Field 2. This is a quick way to compare the Field 1 VITS and Field 2
VITS or to check for the correct insertion of the half line at the end of Field 1.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...