Horizontal Controls
2
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
45
Notice how the time/div information in the status line changes.
The
∇
symbol at the top of the display indicates the time reference point.
The horizontal scale knob works (in the Normal time mode) while
acquisitions are running or when they are stopped. When running,
adjusting the horizontal scale knob changes the sample rate. When
stopped, adjusting the horizontal scale knob lets you zoom into acquired
data. See
"Panning and Zooming Single or Stopped Acquisitions"
on
page 46.
Note that the horizontal scale knob has a different purpose in the Zoom
display. See
"To display the zoomed time base"
on page 50.
To adjust the horizontal delay (position)
1
Turn the horizontal delay (position) knob (
).
The trigger point moves horizontally, pausing at 0.00 s (mimicking a
mechanical detent), and the delay value is displayed in the status line.
Changing the delay time moves the trigger point (solid inverted triangle)
horizontally and indicates how far it is from the time reference point
(hollow inverted triangle
∇
). These reference points are indicated along
the top of the display grid.
Figure 2
shows the trigger point with the delay time set to 200 µs. The
delay time number tells you how far the time reference point is located
from the trigger point. When delay time is set to zero, the delay time
indicator overlays the time reference indicator.
All events displayed left of the trigger point happened before the trigger
occurred. These events are called pre- trigger information, and they show
events that led up to the trigger point.
Everything to the right of the trigger point is called post- trigger
information. The amount of delay range (pre- trigger and post- trigger
information) available depends on the time/div selected and memory
depth.
The horizontal position knob works (in the Normal time mode) while
acquisitions are running or when they are stopped. When running,
adjusting the horizontal scale knob changes the sample rate. When
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...