When
Fine
adjustment is selected, you can change the channel's vertical
sensitivity in smaller increments. The channel sensitivity remains fully
calibrated when
Fine
is on.
The vertical scale value is displayed in the status line at the top of the
display.
When
Fine
is turned off, turning the volts/division knob changes the
channel sensitivity in a 1- 2- 5 step sequence.
To invert a waveform
1
Press the desired channel key.
2
In the Channel Menu, press the
Invert
softkey to invert the selected
channel.
When
Invert
is selected, the voltage values of the displayed waveform are
inverted.
Invert affects how a channel is displayed. However, when using basic
triggers, the oscilloscope attempts to maintain the same trigger point by
changing trigger settings.
Inverting a channel also changes the result of any math function selected
in the Waveform Math Menu or any measurement.
Setting Analog Channel Probe Options
1
Press the probe's associated channel key.
2
In the Channel Menu, press the
Probe
softkey to display the Channel
Probe Menu.
This menu lets you select additional probe parameters such as
attenuation factor and units of measurement for the connected probe.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...