Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
8
Labels
Turning the Entry knob selects a character to enter into the highlighted
position shown in the "New label =" line above the softkeys and in the
Spell
softkey. Labels can be up to ten characters in length.
4
Press the
Enter
softkey to enter the selected character and to go to the
next character position.
5
You may position the highlight on any character in the label name by
successively pressing the
Enter
softkey.
6
To delete a character from the label, press the
Enter
softkey until the
letter you want to delete is highlighted, then press the
Delete Character
softkey.
7
When you are done entering characters for the label, press the
Apply New Label
softkey to assign the label to the selected channel.
When you define a new label, it is added to the nonvolatile label list.
Label Assignment
Auto-Increment
When you assign a label ending in a digit, such as ADDR0 or DATA0, the
oscilloscope automatically increments the digit and displays the modified
label in the "New label" field after you press the
Apply New Label
softkey.
Therefore, you only need to select a new channel and press the
Apply New
Label
softkey again to assign the label to the channel. Only the original
label is saved in the label list. This feature makes it easier to assign
successive labels to numbered control lines and data bus lines.
To load a list of labels from a text file you create
It may be convenient to create a list of labels using a text editor, then
load the label list into the oscilloscope. This lets you type on a keyboard
rather than edit the label list using the oscilloscope's controls.
You can create a list of up to 75 labels and load it into the oscilloscope.
Labels are added to the beginning of the list. If more than 75 labels are
loaded, only the first 75 are stored.
To load labels from a text file into the oscilloscope:
N O T E
You can use a connected USB keyboard instead of using the Spell (and other) character
editing softkeys.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...