![Agilent Technologies 2000 X User Manual Download Page 210](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_user-manual_2868572210.webp)
210
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
16
Waveform Generator
To add noise to the waveform generator output
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Add Noise
softkey
and turn the Entry knob to select the amount of white noise to add to
the waveform generator output.
Note that adding noise affects edge triggering on the waveform generator
source (see
"Edge Trigger"
on page 115) as well as the waveform generator
sync pulse output signal (which can be sent to TRIG OUT, see
"Setting the
Rear Panel TRIG OUT Source"
on page 246). This is because the trigger
comparator is located after the noise source.
To add modulation to the waveform generator output
Modulation is where an original carrier signal is modified according to the
amplitude of a second modulating signal. The modulation type (AM, FM, or
FSK) specifies how the carrier signal is modified.
To enable and set up modulation for the waveform generator output:
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
Softkey (logic levels)
Low level
High level, 50 ohm
expected output load
High level, high-Z
expected output load
TTL
0 V
+2.5 V (TTL compatible)
+5 V
CMOS (5.0V)
0 V
Not Available
+5 V
CMOS (3.3V)
0 V
+2.5 V (CMOS
compatible)
+3.3 V
CMOS (2.5V)
0 V
+2.5 V
+2.5 V
ECL
-1.7 V
-0.8 V (ECL compatible)
-0.9 V
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...