Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10
Trigger Mode/Coupling
To select the Auto or Normal trigger mode
When the oscilloscope is running, the trigger mode tells the oscilloscope
what to do when triggers are not occurring.
In the
Auto
trigger mode (the default setting), if the specified trigger
conditions are not found, triggers are forced and acquisitions are made so
that signal activity is displayed on the oscilloscope.
In the
Normal
trigger mode, triggers and acquisitions only occur when the
specified trigger conditions are found.
To select the trigger mode:
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Mode
softkey; then
select either
Auto
or
Normal
.
See the following
"When to Use Auto Trigger Mode"
on page 137 and
"When to Use Normal Trigger Mode"
on page 137 descriptions.
You can also configure the
[Quick Action]
key to toggle between the Auto
and Normal trigger modes. See
"Configuring the [Quick Action] Key"
on
page 253.
Triggering and
the Pre- and
Post-Trigger
Buffers
After the oscilloscope starts running (after pressing
[Run]
or
[Single]
or
changing the trigger condition), the oscilloscope first fills the pre- trigger
buffer. Then, after the pre- trigger buffer is filled, the oscilloscope starts
searching for a trigger, and sampled data continues to flow data through
the pre- trigger buffer in a first- in first- out (FIFO) manner.
When a trigger is found, the pre- trigger buffer contains the events that
occurred just before the trigger. Then, the oscilloscope fills the post- trigger
buffer and displays the acquisition memory. If the acquisition was
initiated by
[Run/Stop]
, the process repeats. If the acquisition was initiated
by pressing
[Single]
, the acquisition stops (and you can Pan and Zoom the
waveform).
In either Auto or Normal trigger mode, a trigger may be missed if the
event occurs while the pre- trigger buffer is being filled. This may be more
likely, for example, when the horizontal scale knob is set to a slow
time/div setting, such as 500 ms/div.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...