![Agilent Technologies 2000 X User Manual Download Page 192](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_user-manual_2868572192.webp)
192
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
14
Mask Testing
Run Until
The Run Until softkey lets you specify a condition on which to terminate
testing.
• Forever — The oscilloscope runs continuously. However, if an error occurs
the action specified using the On Error softkey will occur.
• Minimum # of Tests — Choose this option and then use the # of Tests
softkey to select the number of times the oscilloscope will trigger, display
the waveform(s), and compare them to the mask. The oscilloscope will
stop after the specified number of tests have been completed. The
specified minimum number of tests may be exceeded. If an error occurs the
action specified using the On Error softkey will occur. The actual number of
tests completed is displayed above the softkeys.
• Minimum Time — Choose this option and then use the Test Time softkey
to select how long the oscilloscope will run. When the selected time has
passed the oscilloscope will stop. The specified time may be exceeded. If
an error occurs the action specified using the On Error softkey will occur.
The actual test time is displayed above the softkeys.
• Minimum Sigma — Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the
oscilloscope will perform the action specified by the On Error softkey.)
Note that this is a test sigma (the max achievable process sigma,
assuming no defects, for a certain number of tested waveforms) as
opposed to a process sigma (which is tied to the amount of failures per
test). The sigma value may exceed the selected value when a small sigma
value is chosen. The actual sigma is displayed.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...