To manually modify a mask file
You can manually modify a mask file that you created using the Automask
function.
1
Follow the steps 1- 7 in
"To create a mask from a "golden" waveform
(Automask)"
on page 189. Do not clear the mask after creating it.
2
Attach a USB mass storage device to the oscilloscope.
3
Press the
[Save/Recall]
key.
4
Press the
Save
softkey.
5
Press the
Format
softkey and select
Mask
.
6
Press the second softkey and select a destination folder on your USB
mass storage device.
7
Press the
Press to Save
softkey. This creates an ASCII text file that
describes the mask.
8
Remove the USB mass storage device and connect it to a PC.
9
Open the .msk file your created using a text editor (such as Wordpad).
10
Edit, save, and close the file.
The mask file contains the following sections:
•
Mask File Identifier.
•
Mask Title.
•
Mask Violation Regions.
Reset Statistics
Note that statistics are also reset when:
• Mask Test is switched on after being switched off.
• Clear Mask softkey is pressed.
• An Automask is created.
Additionally, the accumulated time counter is reset whenever the
oscilloscope is run after the acquisition was stopped.
Transparent
Enable Transparent mode to write measurement values and statistics on the
screen without a background. Disable the Transparent mode to show them
with a gray background. The Transparent setting affects mask test statistics,
measurement statistics, and reference waveform information display.
Clear Display
Clears acquisition data from the oscilloscope display.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...