Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
17
Save/Recall (Setups, Screens, Data)
•
Whether acquisitions are running. When stopped, data comes from the
raw acquisition record. When running, data comes from the smaller
measurement record.
•
Whether the oscilloscope was stopped using
[Stop]
or
[Single]
. Running
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
•
Whether only one channel of a pair is turned on. (Channels 1 and 2 are
one pair, channels 3 and 4 are the other.) Acquisition memory is
divided among the channels in a pair.
•
Whether reference waveforms are on. Displayed reference waveforms
consume acquisition memory.
•
Whether digital channels are on. Displayed digital channels consume
acquisition memory.
•
Whether segmented memory is on. Acquisition memory is divided by
the number of segments.
•
The horizontal time/div (sweep speed) setting. At faster settings, fewer
data points appear on the display.
•
When saving to a CSV format file, the maximum number of data points
is 50,000.
When necessary, the Length control performs a "1 of n" decimation of the
data . For example: if the
Length
is set to 1000, and you are displaying a
record that is 5000 data points in length, four of each five data points will
be decimated, creating an output file 1000 data points in length.
When saving waveform data, the save times depend on the chosen format:
See Also
•
"Binary Data (.bin) Format"
on page 281
•
"CSV and ASCII XY files"
on page 287
•
"Minimum and Maximum Values in CSV Files"
on page 288
Data File Format
Save Times
BIN, ALB
fastest
ASCII XY
medium
CSV
slowest
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...