Trigger Mode/Coupling
10
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
139
Note that Trigger Coupling is independent of Channel Coupling (see
"To
specify channel coupling"
on page 57).
To enable or disable trigger noise rejection
Noise Rej adds additional hysteresis to the trigger circuitry. By increasing
the trigger hysteresis band, you reduce the possibility of triggering on
noise. However, this also decreases the trigger sensitivity so that a slightly
larger signal is required to trigger the oscilloscope.
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Noise Rej
softkey to
enable or disable.
To enable or disable trigger HF Reject
HF Reject adds a 50 kHz low- pass filter in the trigger path to remove high
frequency components from the trigger waveform.
You can use HF Reject to remove high- frequency noise, such as AM or FM
broadcast stations or noise from fast system clocks, from the trigger path.
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
HF Reject
softkey to
enable or disable.
Pass
Band
50 kHz
DC
0 dB
-3 dB
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 18: ...18 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 42: ...42 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 76: ...76 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 100: ...100 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 106: ...106 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Display Settings ...
Page 134: ...134 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 9 Triggers ...
Page 142: ...142 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Trigger Mode Coupling ...
Page 168: ...168 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 12 Cursors ...
Page 204: ...204 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Digital Voltmeter ...
Page 216: ...216 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 236: ...236 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Print Screens ...
Page 272: ...272 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Web Interface ...
Page 290: ...290 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Reference ...