Troubleshooting
The warming unit software routinely checks unusual operating conditions. If such a condition occurs, the system may trigger a fault condition and alarm. The warming unit can be
reset by disconnecting the warming unit from the main power source and allowing 30-60 seconds for the system to reset. If the fault code condition reoccurs, disconnect again
and wait 5 minutes for the system to reset. If the fault condition reoccurs upon connection to the main power source, the warming unit may have a hard-fault or permanent error.
If the fault does not reoccur upon connection to the main power source, the warming unit may have a cautionary fault.
Fault Code
Fault Condition
Possible Cause
Troubleshooting
01
Control sensor output
is high
1. Sensor board failure.
2. Control board failure.
3. Internal wire harness failure.
1. Check for moisture on sensor.
2. Perform the 43C test in the
System Test
option of the
Diagnostic Menu
.
Verify the end of hose temperature is 43°C ± 3°C using the Model
22110 temperature test unit and calibrated thermocouple meter.
02
Control sensor measures ≥
47°C for 30 seconds
1. Control board failure.
2. AC board failure (TRIAC).
3. Sensor board failure.
4. Internal wire harness failure.
1. Perform Operating Temperature calibration.
2. Measure the hose end temperature using the Model 22110 temperature
test unit and calibrated thermocouple meter while running in
Ambient
mode. The measured temperature should be ±3°C of the ambient
room temperature.
03
Monitor sensor output
is high
1. Sensor board failure.
2. Control board failure.
3. Internal wire harness failure.
1. Check for moisture on sensor.
2. Perform the 43C test in the System Check option of the
Diagnostic
Menu
. Verify the end of hose temperature is 43°C ± 3°C using the Model
22110 temperature test unit and calibrated thermocouple meter.
04
Software detects an
over-temperature
circuit cutout
1. Control board failure.
2. Sensor board failure.
3. Internal wire harness failure.
1. Perform Operating Temperature calibration.
2. Measure the hose end temperature using the Model 22110 temperature
test unit and calibrated thermocouple meter while running in
Ambient
mode. The measured temperature should be ±3°C of the ambient
room temperature.
50
Software detects any button
pressed for 5 seconds.
1. User action.
2. User interface circuit board failure.
1. Reset microprocessor by disconnecting the warming unit from the main
power source and allowing 30-60 seconds for the system to reset.
51
The blower is running at ≤
2000 RPM for 1 minute, or
there is no output to the
control board for 3 seconds
1. Blower failure.
2. Control board failure.
3. Internal wire harness failure.
1. Test blower using the
System Info
screen in
Diagnostic
mode.
52
Watchdog timer failed
1. Control board failure.
1. Reset microprocessor by disconnecting the warming unit from the main
power source and allowing 30-60 seconds for the system to reset.
70
The control sensor measures
the temperature to be +3°C
greater than the set point for
3 minutes.
1. Control board failure.
2. Air flow occlusion.
1. Check hose for obstructions.
2. Check the air intake on the bottom of the unit for obstructions.
3. Check for blocked filter.
4. Test all 3 heater wattages using the
System Test
screen in the
Diagnostic Menu
.
71
The control sensor measures
the temperature to be -3°C
less than the set point for
3 minutes.
1. Air flow occlusion.
2. Heater failure.
3. Control board failure.
4. Internal wire harness failure.
1. Check for hose obstructions.
2. Check the air intake on the bottom of the unit for obstructions.
3. Check for blocked filter.
4. Test all 3 heater wattages using the
System Test
screen in
Diagnostic
mode.
72
The outlet temperature
does not increase at
expected rate
1. Air flow occlusion.
2. Heater failure.
3. Control board failure.
4. Internal wire harness failure.
1. Check for hose obstructions.
2. Check the air intake on the bottom of the unit for obstructions.
3. Check for blocked filter.
4. Test all 3 heater wattages using the
System Test
screen in
Diagnostic
mode.
73
Control and monitor sensors
differ by more than 3°C for
30 seconds.
1. Sensor board failure.
2. Internal wire harness failure.
3. Control board failure.
1. Check for moisture on sensor.
2. Perform the 43C test in the System Check option of the Diagnostic
Menu. Verify the end of hose temperature is 43°C ± 3°C using the Model
22110 temperature test unit and calibrated thermocouple meter.
Technical Support and Order Placement
When you call for Technical Support
We will need to know the serial number of your Bair Hugger warming unit when you
call us. The serial number label is located on the back of the warming unit.
USA:
• Call 3M Technical Support at 800-733-7775
Outside of the USA:
• Contact your local 3M representative
In-Warranty Repair and Exchange
USA
• Call 3M technical support at 800-733-7775 if your warming unit requires
factory service.
• Provide the serial number located on the back of the warming unit.
• Technical Support will give you a Return Authorization (RA) number. Use this RA
number on all correspondence. Your Technical Support representative will also
send a shipping carton to you at no charge, if needed.
Outside of the USA
• Contact your local distributor concerning in-warranty repair and exchange.
Provide the serial number located on the back of the Bair Hugger warming unit
when you call.
Summary of Contents for Bair Hugger 675
Page 2: ......
Page 4: ......
Page 6: ......
Page 19: ...English 34 8720 2487 1 15 Maintenance Log Date Maintenance Action Performed Hour Meter ...
Page 20: ......
Page 22: ......
Page 36: ...Journal d entretien Date Procédure d entretien effectuée Compteur d heures ...
Page 38: ......
Page 54: ......
Page 68: ...Registro di manutenzione Data Azione di manutenzione eseguita Contaore ...
Page 70: ......
Page 86: ......
Page 99: ...Nederlands 34 8720 2487 1 95 Onderhoudslogboek Datum Uitgevoerde onderhoudshandeling Tijdstip ...
Page 100: ......
Page 102: ......
Page 115: ...Svenska 34 8720 2487 1 111 Skötseljournal Datum Skötselåtgärd utförd Timmätare ...
Page 116: ......
Page 118: ......
Page 131: ...Dansk 34 8720 2487 1 127 Vedligeholdelslog Dato Udført vedligeholdeseshandling Timemåler ...
Page 132: ......
Page 134: ......
Page 147: ...Norsk 34 8720 2487 1 143 Vedlikeholdslogg Dato Type vedlikehold utført Timeteller ...
Page 148: ......
Page 150: ......
Page 164: ...Huoltoloki Päiväys Suoritettu huoltotoimenpide Tuntimittari ...
Page 166: ......
Page 180: ......
Page 182: ......
Page 200: ......
Page 214: ......
Page 216: ......
Page 230: ......
Page 232: ......
Page 245: ...Čeština 34 8720 2487 1 241 Protokol o údržbě Datum Provedená údržba Měřič hodin ...
Page 246: ......
Page 248: ......
Page 262: ......
Page 264: ......
Page 282: ......
Page 295: ...Turkce 34 8720 2487 1 291 Bakım Günlüğü Tarih Gerçekleştirilen Bakım Faaliyeti Saat Sayacı ...
Page 296: ......
Page 298: ......
Page 311: ...傈劤铂 34 8720 2487 1 307 ًٝ شذ ٝ ؚٗأ 傈 㹋倵 ً ٝ شذ ٝ أ 噟 儗 鎘 ...
Page 312: ......
Page 314: ......
Page 329: ...ƶǻƴLjǞǬ 34 8720 2487 1 325 ƶDzƲǻǒǬ ǫƿNj ƲǝƲnjǬ džǝ GLJǻǤdzƸ ǮƸ JLJǬ ƶDzƲǻǒǬ Ljƾ ǃǺ1ƲƹǬ ...
Page 330: ......
Page 331: ......