Safety measures
S100 / OPMI pico
Version 9.0
Page 18
G-30-1781-en
Patient-related factors:
–
The general condition of a patient's health may contribute to the risk of
injury.
–
The skin type may also play a major role for the risk of injury.
–
Certain medications also affect the sensitivity to light.
–
The interaction of heat and antimicrobial substances in incision foils may
lead to an increased reaction of the patient to these substances.
Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, ZEISS cannot provide guidance on acceptable
illumination intensities and exposure durations. However, this system has
several features that can help the user to reduce the risk of burns.
–
The start value of the illumination intensity can be set to a low value.
–
The risk of burns can be reduced by constantly by irrigating the illuminated
surgical field and by keeping it moist.
–
If used, surgical drapes should also be remoistened at regular intervals in
order to prevent heat from accumulating underneath the surgical drape.
–
Switch off the light when the microscope is not in use, and make sure that
it is not aimed at unprotected bare skin.
Please note that in most published cases of burns the skin around the incision
was affected. The most important measure for avoiding burns is to cover the
peripheral area with wet, sterile gauze. The area of the incision should be
constantly moistened.
Concluding remarks
ZEISS recommends that the illumination of the surgical field be kept to the
necessary minimum, for the patient's safety and for a good microscopic
image.
Please refer to the relevant warnings and precautions.
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...