Care and Maintenance
S100 / OPMI pico
Version 9.0
Page 230
G-30-1781-en
water plus a dash of household dish-washing liquid.
Sterilization
The asepsis sets available from ZEISS contain rubber caps and hand grips that
can be sterilized in autoclaves. For detailed information on sterilization, please
see the enclosed instructions "Preparation of resterilizable products" for the
respective asepsis set.
You can also use disposable sterile drapes to cover the system.
•
Loosely attach the sterile drapes so that there is sufficient space for the
microscope supporting stand and the surgical microscope to move.
•
Do not cover the ventilation slots in order to ensure that the lamps are
able to cool sufficiently and to avoid any loss in lamp functionality.
Disinfection
The maximum application concentrations are:
–
For alcohol (tested with 2 propanol): 60%
–
For aldehyde (tested with glutaraldehyde): 2%
–
For quaternary compounds (tested with DDAC): 0.2%
NOTE
This could damage the paint on the device.
•
Use a disinfectant based on aldehyde or alcohol. The addition of
quaternary compounds is acceptable.
To avoid damaging the surfaces, do not use any disinfecting components
other than those specified below.
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...