What to do in the event of malfunctions
S100 / OPMI pico
Version 9.0
Page 210
G-30-1781-en
Lamp in light source
remains dark and fan is
running
Knob for brightness control
set to left-hand stop.
Turn knob clockwise.
Inclination switch
deactivated.
Move suspension arm of
surgical microscope down
from its standby position.
–
Inclination switch out of
adjustment or defective
Notify Service.
–
Lamp module has no
contact.
Insert lamp module to the
stop.
–
In a halogen light source:
Halogen lamp defective.
Switch to backup lamp.
In a halogen light source:
Ceramic base does not have
proper contact with
halogen lamp.
Plug ceramic base firmly
onto contacts of halogen
lamp.
In a xenon light source:
Xenon lamp defective.
Switch to backup lamp.
In a LED light source:
LED light source defective.
Notify Service.
–
Defective stand electronics.
Illuminate surgical area with
an OR illuminator.
Notify Service.
–
Lamp goes constantly off
and on again during
operation.
Ventilation openings are
covered or dirty.
Ventilation openings must
be clear; clean them if
necessary.
–
Defective fan.
Illuminate surgical area
with an OR illuminator.
Notify Service.
–
Failure of suspension
electronics.
Illuminate surgical area
with an OR illuminator.
Notify Service.
–
Defective magnification
changer.
-
Notify Service.
–
Filters in illumination beam
path are defective or cannot
be exchanged.
-
Notify Service.
–
Fault
Possible reason
Corrective measure
Cross
reference
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...