Device overview
S100 / OPMI pico
Version 9.0
Page 72
G-30-1781-en
xenon light source
1
Ventilation grid
2
Lamp module
3
Manual switching to the backup lamp
The xenon light source has a lamp module containing two xenon lamps.
The second lamp is used as backup lamp. Manually swivel in the backup
lamp into position (Page 217) if the maximum service life of 500 hours has
been reached or the first lamp has failed (see also maintenance intervals
Page 222).
If the backup lamp is being used, keep a new lamp module ready for
exchange. (Page 224).
4
Indicator: backup lamp is in use
If the red segment (4) in the button (3) is lit, the backup lamp is operated.
5
Opening the lamp module
When you press this button, the slide-in lamp module is partially ejected.
6
Brightness setting
If the light source is switched on, the brightness can be infinitely adjusted
by turning the brightness control button for the fiber illumination.
CAUTION
Limited service life of the xenon lamp!
If used longer than the indicated service life of 500 hours the xenon lamp
may fail or burst.
Dropping fragments may contaminate the sterile area. Risk of infection.
•
Manually swivel in the backup lamp (Page 217) or replace the lamp
module in a timely manner (see Page 226).
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...