Care and Maintenance
S100 / OPMI pico
Version 9.0
Page 222
G-30-1781-en
Device maintenance
Electromagnetic
compatibility (EMC)
No regular inspections or maintenance procedures are required in order to
maintain electromagnetic compatibility (EMC).
Maintenance intervals
Regular preventive maintenance is required in order to ensure safe and correct
operation of the device as well as a long service life.
The following table lists the relevant maintenance activities and shows at
which maximum intervals they must be performed.
The maintenance activities that you are able to perform yourself are described
below.
Any other maintenance activities that is not described here in detail requires
device-specific knowledge.
Please contact your local ZEISS Service or authorized partner to arrange for
the timely performance of these maintenance activities.
Interval
Maintenance tasks
Max. 12 weeks
Exchange xenon lamp if it is operated for 8 hours / day on 5 days / week.
This corresponds to the service life of a xenon lamp of 500 h.
Max. 25 weeks
Exchange xenon lamp if it is operated for 4 hours / day on 5 days / week.
This corresponds to the service life of a xenon lamp of 500 h.
Every 2 years
Visual inspection and function test of the following modules:
–
S100 suspension system
–
Illumination and power supply unit
–
Surgical microscope with tube
–
Light guide
–
Power cord
–
Connected accessories
–
Secure seating of OPMI suspension
–
Securing screws are available and tightened
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...