Safety measures
S100 / OPMI pico
Version 9.0
Page 20
G-30-1781-en
xenon light source
–
UV / IR heat protection filter
The filter protects against UV and IR radiation.
–
Temperature monitoring
A sensor constantly monitors the lamp temperature.
1
Switching to the backup lamp
The xenon light source has a lamp module containing two xenon lamps.
The second lamp is used as backup lamp. Manually swivel in the backup
lamp into position (Page 217), if the maximum service ;life of 500 hours
is reached (see chapter "Maintenance intervals" on Page 222). When the
red segment (2) in the knob (1) lights up, the backup lamp is in use.
Keep the lamp module ready if you work with the backup lamp.
If the backup lamp is operated, replace the lamp module in a timely
manner (see Page 224) before it is used beyond its maximum service life
of 500 hours (see chapter "Maintenance intervals on Page 222).
Fig. 2: Safety system xenon
light source
CAUTION
Limited service life of the xenon lamp!
If used longer than the indicated service life of 500 hours the xenon lamp
may fail or burst.
Dropping fragments may contaminate the sterile area. Risk of infection.
•
Change the lamp module on time.
1
3
2
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...