Adjusting the surgical microscope
Notes on adjustment
–
To permit fast setup of the microscope, it is advisable to store the pupillary
distances and prescriptions of the different users so that they can be easily
preset during system preparation.
–
Do not wear any bifocal or progressive lenses for the adjustment
procedure and subsequent work with the microscope.
Such lenses make it impossible to set the diopters correctly and result
in unsatisfactory imaging.
–
Optimum image clarity in the eyepieces and at the optical port for
documentation equipment (photo, video) can only be achieved through
careful focusing and precisely adjusting the binocular tube.
Adjusting the binocular tube
Adjusting the
interpupillary distance
•
Position the microscope vertically above a flat object, e.g. a piece of paper
with writing on it.
•
Bring the eyepieces on the binocular tube to eye distance, so that the two
eyepiece images (the object and the periphery of the field of view) merge
to produce a single image.
Adjusting the eyepieces
The following procedure must be performed separately for each eyepiece in
the order described.
•
Set to the smallest magnification on the microscope.
•
Set the diopter adjustment ring at the eyepiece to zero position.
•
Look through the eyepiece and move the microscope to focus the image.
•
Set the microscope to the highest magnification and correct the fine
focusing until the image is defined sharply.
•
Now set to the smallest magnification once again without changing the
working distance.
•
Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
•
Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
•
Repeat the entire procedure for the second eyepiece.
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...