What to do in the event of malfunctions
S100 / OPMI pico
Version 9.0
Page 208
G-30-1781-en
Troubleshooting
For your safety
•
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions", attach a sign to the system
stating it is out of order and contact our Service staff.
System malfunctions
Faults on S100 stands with video (option)
Fault
Possible reason
Corrective measure
Cross
reference
No function at all
Suspension system
turned off.
Switch on suspension
system or plug in power
cord.
Power plug of stand not
inserted.
Plug in the power plug.
Mains connection defective.
Replace the power cord.
–
Power failure
Contact in-house
electrician.
–
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...