Instructions for Use
5 Daily startup
OPMI LUMERA 700
5.2 Adjusting the surgical and assistant's microscopes
G-30-1673-en - 14.0 - 2018-07-12
115 / 236
5.2.4 Adjusting microscope tilt
The manual tilting device can be used to orient the optical axis of
the surgical microscope vertically to the patient's eye.
CAUTION!
Risk of injury caused by assistant's microscope swiveling in!
If you tilt the surgical microscope in the horizontal viewing
direction and the assistant’s microscope is not locked in place, the
assistant's microscope may swivel in.
u
Position the assistant's microscope and lock it before beginning
surgery.
CAUTION!
Risk of injury caused by fundus viewing system swiveling in!
If the RESIGHT 500 or RESIGHT 700 fundus viewing system is
mounted on the underside of the microscope, and you steeply tilt
the microscope, the RESIGHT fundus viewing system may swivel in.
u
Remove the fundus viewing system before tilting the
microscope at a steep angle.
Action
1. Tilt the surgical microscope towards you: rotate the [OPMI tilt]
knob clockwise.
2. Tilt the surgical microscope away from you: rotate the [OPMI
tilt] knob counter clockwise.
Result
ü
When the bar on the surgical microscope and that of the
support arm are perpendicular to each other, the surgical
microscope is in the perpendicular position relative to the
patient's eye.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...