5 Daily startup
Instructions for Use
OPMI LUMERA 700
5.1 Outfitting the surgical and assistant's microscopes with accessories
110 / 236
G-30-1673-en - 14.0 - 2018-07-12
5.1.3 Connecting RESIGHT 700
The RESIGHT 700 internal focus is controlled by the electrical
connection. If RESIGHT 700 is connected along with the
Invertertube E, both accessories are synchronized and the field of
view is always displayed in the correct orientation. This also
includes the image inversion of an integrated HD camera.
Prerequisite
þ
RESIGHT 700 is attached.
Action
1. Insert the RESIGHT 700 connection cable into the surgical
microscope's RESIGHT 700 connector.
5.1.4 Connecting auxiliary illumination
On devices with dual light sources, an auxiliary illumination system
can be connected by means of a second light guide.
CAUTION!
Risk of injury caused by suspended light guide!
A suspended light guide may cause retina damage or slight burns
to the patient.
u
Always turn off the light source for the light guide that is not in
use.
u
Push the light guide that isn't required into a free space on the
cable holder so that it is pointing away from the patient.
u
Remove the protective cap on the unused light guide. This will
prevent melting of the protective cap if the light source is
accidentally turned on.
Material
• S light guide, 2.5 m
Prerequisite
þ
The auxiliary illumination system is attached.
þ
The device is switched off.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...