2 Safety notes
OPMI LUMERA 700
2.6 Requirements for operation
G-30-1673-en - 14.0 - 2018-07-12
21 / 236
Hazards arising from moving components
The following safety measures should be observed in order to
prevent injuries or damage:
u
Make certain to compensate for any added weight to the
device; this will enable the surgical microscope to maintain its
balance in all positions of the working range.
u
Limit the stroke of the suspension arm to prevent any contact
with the patient should the surgical microscope be lowered
accidentally.
u
Check the surgical microscope for sufficient freedom of
movement.
u
Make sure there is sufficient space available for focus
positioning. The surgical microscope must not come into
contact with the patient.
u
Secure the device into position using the locking tabs on the
stand base to prevent the device from rolling away
unexpectedly.
u
When using a fundus viewing system: make sure that the space
available for movement is greater than the downward travel
path of the microscope.
u
When using an assistant's microscope: adjust the assistant’s
microscope before use and make sure it is properly locked in
position. If you tilt the surgical microscope in the horizontal
viewing direction and the assistant’s microscope is not locked
in place, the assistant's microscope may swivel in.
Hazards arising from an attached fundus viewing system
Incorrect handling of a fundus viewing system mounted on the
underside of the microscope, or activation of the fast focus, may
cause injury to the patient’s eye.
u
Before using a fundus viewing system, make sure that the
space available for movement is greater than the downward
travel path of the microscope.
u
Perform a focus reset.
Hazards arising from disruptions to the 14-function foot
control panel
Inadequate power supply to the 14-function foot control panel may
cause the device to malfunction.
u
Make sure that the batteries are fully charged and that the
"Battery" status display is not blinking.
Hazards arising from unknown software settings
The following safety measures should be observed in order to
prevent injuries or damage:
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...