Instructions for Use
10 Troubleshooting
OPMI LUMERA 700
10.4 Troubleshooting work
G-30-1673-en - 14.0 - 2018-07-12
191 / 236
10.4.2 Operating the device in manual mode
1
2
3
Figure 78: Operating the device in manual mode
1
Zoom adjustment knob
2
[SCI illumination] rotary knob
3
[Integrated slit illumination] rotary knob
Prerequisite
þ
Manual mode is activated.
Action
1. Position the surgical microscope. Move the suspension arm in
the desired direction.
2. Focus the surgical microscope. To do this, move the suspension
arm up and down.
3. Set the magnification on the manual zoom adjustment knob.
4. Set the SCI illumination. To do this, rotate the [SCI illumination]
rotary knob to the following positions:
ð
Left position: light emitted generates a red reflex.
ð
Right position: light emitted illuminates the entire field of
view
5. Set the integrated slit illumination. To do this, rotate the
[Integrated slit illumination] rotary knob to the following
positions:
ð
Left position: slit illuminator from the left
ð
Center position: standard OPMI illumination
ð
Right position: slit illuminator from the right
Tip: In manual mode, you can switch off the light sources for light
guides which you no longer need. To do this, press the [Open lamp
module] button on the lamp housing.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...