Instructions for Use
3 Description of the device
OPMI LUMERA 700
3.3 Control elements and displays
G-30-1673-en - 14.0 - 2018-07-12
53 / 236
3.3.5 Overhead display
1
3
4
5
7
6
2
Figure 32: Displayable elements on the overhead display
Item
Symbol
Name
Explanation
1
"CAUTION"
Displays a message on the control
panel. Once the message is
canceled by the user, the message
disappears on the control panel
and on the overhead display.
2
R
Surgery
profiles
Empty circle: displays created
surgery profiles. One circle is
displayed for each surgery profile
which has been configured for the
current user (max 5).
Filled circle: indicates the active
surgery profile.
R: a surgery profile with RESIGHT
profile is active
3
Light source
light
intensity
Displays the light intensity from
5% to 100%.
4
Video status
REC OFF: no video recording.
REC ON: video recording is
running.
REC ON/REC OFF: no integrated
HD image and video recording
available.
5
Focus
position
Up arrow: focus 40 mm above
initial position.
Center arrow: focus at initial
position.
Down arrow: focus 30 mm below
initial position.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...