4 Installation
Instructions for Use
OPMI LUMERA 700
4.3 Connections
98 / 236
G-30-1673-en - 14.0 - 2018-07-12
4.3.4 Setting up a network connection with integrated
CALLISTO eye Panel PC
The device can be connected either directly (point to point) or
indirectly via a switch to an integrated CALLISTO eye Panel PC.
4.3.4.1 Setting up a network connection via a switch
3
2
1
Figure 61: Setting up a network connection via a switch
1
LAN 2 network connection
2
LAN 1 network connection
3
Switch
CAUTION!
Risk of injury due to electrical current!
In the event of a defect in the IT network, the network cable plug
can carry dangerous voltage.
u
Check to make sure that the requirements [
90] for connecting
the device to an IT network are met.
Material
• 2x network cable, 10m
Action
1. Switch the device off.
2. Connect the LAN 1 and LAN 2 network connectors to the
hospital IT network. Cable: 2x network cable, 10 m
3. Switch the device on.