4 Installation
Instructions for Use
OPMI LUMERA 700
4.2 Preparing the installation
90 / 236
G-30-1673-en - 14.0 - 2018-07-12
4.2 Preparing the installation
4.2.1 Requirements for connecting the device to an IT
network
The device can be connected to an existing IT network. This
enables:
• Video and image data to be saved to a shared directory.
• OPMI LUMERA 700 to be connected to CALLISTO eye. This
function enables the configuration / remote control of various
functions of the surgical microscope via the Ethernet interface.
CAUTION!
Risk of damage caused by viruses!
Connecting the device to an existing IT network or connecting a
USB storage medium to the device may infect the device with
viruses! This may lead to limited functionality or malfunctions in the
device.
u
Keep the IT network or USB storage medium free of viruses.
The institution operating the software is responsible for the
safety of the IT network.
CAUTION!
Unknown risks due to network integration!
Integration of the device in an existing IT network which includes
other ME devices may lead to unknown risks for patients, operators
or other persons.
u
The operator of the device must determine, analyze, evaluate
and control these risks before the device is integrated in the IT
network.
u
The following changes to the IT network can result in new risks
and may therefore necessitate additional analyses:
- Changing the IT network configuration
- Attaching auxiliary elements to the IT network
- Removing elements from the IT network
- Updating devices which are connected to the IT network
CAUTION!
Risk of electric shock!
If the IT network does not provide the features required to
integrate the device into the IT network (listed below), the
following hazards may arise:
Risk of electrical shock if the user touches a live part of the casing
or a patient.
Increased leakage current on the power outlets.
The following specifications and measures as per IEC 60601 are to
be adhered to when connecting the device to an existing IT
network:
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...