Instructions for Use
3 Description of the device
OPMI LUMERA 700
3.5 Functional description
G-30-1673-en - 14.0 - 2018-07-12
85 / 236
• Manually set the magnification on the microscope's zoom
adjustment knob.
• Manually position the suspension system for focusing and for
moving the system in the XY direction.
Pressing the manual mode button again switches the device back
to working mode.
3.5.11 Assistant's microscope
The device can be equipped with an assistant's microscope. The
assistant’s microscope is rotatable and has two working positions.
These working positions are to the left and right of the main
surgeon and at a 90° angle. The assistant's microscope can be
swiveled out of the 90° position by using the swivel mechanism.
The swivel mechanism lock holds the assistant's microscope in
position when the surgical microscope is tilted horizontally. The
assistant’s microscope features a focusing system and is equipped
with a motorized zoom system or a manual 5x magnification
changer.
No.
Function
Effect
1
Electrical zoom
system
The electrical zoom system works
independently of, or in parallel with, the
surgical microscope zoom feature.
2
Alternative: manual
5x magnification
changer
The manual 5x magnification changer
works independently of the surgical
microscope's zoom feature.
3.5.12 IDIS (Integrated Data Injection System)
The device can be equipped with IDIS. IDIS enables data to be
injected into the device's right eyepiece or CALLISTO eye.
The following device status information can be displayed:
• Main light source brightness
• Auxiliary light source brightness
• Zoom system magnification
• Charge state of the battery for the foot control panel
• Signal strength of the foot control panel radio connection
Recorded OCT images and the visible area on the screen and in the
video can be injected from CALLISTO eye.
If CALLISTO eye is equipped with the "Assistance Functions"
option, additional assistance functions data can be injected.
• K TRACK : visualization of the shape of the cornea in combi-
nation with the keratoscope ring.
• Z ALIGN : an assistance function for aligning toric intraocular
lenses.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...