11 Technical specifications
Instructions for Use
OPMI LUMERA 700
11.9 Guidance and manufacturer´s declaration – electromagnetic immunity
212 / 236
G-30-1673-en - 14.0 - 2018-07-12
WARNING!
Loss of performance caused by HF units!
u
Do not use any portable or mobile HF communication
equipment or transmitters (including peripheral devices such as
antenna cables or external antennas) in the proximity of the
OPMI LUMERA 700 (minimum distance 30 cm). It cannot be
ruled out that the function of the device will be impaired or
that the performance of the device will deteriorate.
u
Do not use cell phones near the OPMI LUMERA 700. They
represent a potential risk for the proper functioning of medical
equipment. Malfunctions may occur, depending on a variety of
local factors. They cannot be predicted and can by no means
be estimated.
u
Please follow the EMC guidelines on the following pages.
NOTE
Disturbances caused by electromagnetic radiation!
The OPMI LUMERA 700 may experience interference from other
devices even if the other devices comply with the emission require-
ments applicable to them as per CISPR.
u
Do not use the OPMI LUMERA 700 when it is located next to,
or stacked on top of, other devices.
u
If it is necessary to operate the device next to, or stacked on
top of, other devices, monitor the OPMI LUMERA 700 to
ensure its normal operation in the arrangement in which it is
used.
u
If you use the OPMI LUMERA 700 together with other devices
that generate high-frequency waves or high voltages, check the
effects that these devices have before using them, and install
the device in such a way so as to minimize disturbances caused
by high-frequency waves.
The following device functions have been defined for the immunity
test:
• The CALLISTO eye user interface is accessible at all times.
• The integrated HD camera's video signal is accessible at all
times.
• Magnification of the surgical field is accessible at all times.
• Illumination of the surgical field is accessible at all times.
• Focusing on the surgical field is accessible at all times.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...