<2. SPECIFICATIONS>
2-3
IM 12E04A02-02E
Maintenance Functions (<MAINT.> mode):
Zero calibration, shift calibration, span calibration, sensitivity correction
calibration, manual start of auto cleaning, manual start of auto zero
calibration, error code indication, error reset and calibration of lamp
control reference value.
Setting Functions:
<PROGRAM.1> mode: Setting of output range, average coef
fi
cient, output value in
failure condition, bi-linear output, high alarm, low alarm and spike
detection function.
<PROGRAM.2> mode: Setting of auto cleaning time and auto zero calibration time, and
selection of enabling/disabling failure detection functions.
Installation Location: Indoors (For outdoor installation, a protection against rain is required
separately.)
Installation:
Detector and converter: Pipe or rack mounting
With sampling system: Anchor bolt mounting
Piping Connection:
With sampling system: Drain: VP40
Others:
VP16
Without sampling system: 25-mm ID hose coupling
Cable Inlet:
5 cable glands (bottom of the converter)
Applicable wire:
6 to 12 mm OD
Wiring:
Power, analog output, digital output, contact output, contact input
and ground (using the ground terminal inside the converter or the
ground terminal on the outside of the case.)
Water Sample:
With sampling system:
Flow rate:
2 to 10 l/min
Pressure:
20 to 500 kPa
Temperature:
0 to 50 °C (Ambient temperature should not exceed 30 °C.)
Without sampling system:
Flow rate:
1.5 to 2 l/min
Temperature:
0 to 50 °C (Ambient temperature should not exceed 30 °C.)
Zero Turbidity Water and Cleaning Water (Tap Water): (With Sampling System)
Turbidity:
2 mg/l or less
Temperature:
0 to 50 °C (Ambient temperature should not exceed 30 °C)
Pressure:
100 to 500 kPa
Flow rate:
Zero turbidity water:2 to 10 l/min
Cleaning water:
3 to 6 l/min
Consumption:
Zero turbidity water:Approx. 380 l/day (at a
fl
ow rate of 2 l/min)
Cleaning water: Approx. 90 l/min (at a
fl
ow rate of 3 l/min)
(when auto cleaning/zero calibration is enabled and time setting is
factory default.)
Power Supply:
100/110 V AC, 50/60 Hz, or 200/220 V AC, 50/60 Hz
Power Consumption:
Detector and converter: 50 VA or less, respectively
With sampling system: 200 VA or less (in full speci
fi
cations)
Automatic Cleaning Function: Water jet cleaning. Cleaning time and interval are user
con
fi
gurable. (When auto cleaning is speci
fi
ed.)
Automatic Zero Calibration Function: Zero point calibration using zero turbidity water (When auto
cleaning/zero calibration is speci
fi
ed.)
3rd Edition : May. 31, 2010-00