<6. PROCEDURES FOR CONVERTER OPERATION>
6-33
IM 12E04A02-02E
The following shows an example of the setting operations for the automatic cleaning period.
Similar operations apply to the other settings.
(Example of display)
1. Select FUNCTION "1" in the <PROGRAM2>
=> 0002.0 (time)
mode. The current setpoint is displayed.
2. Press the [>] key.
=> 0002.0 (time)
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set the digit
=> 0004.0 (time)
to the desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key.
=> 0004.0 (time)
The entire display
fl
ashes once.
If the setpoint in step 4 is beyond the permissible range, "not" is displayed. Reset the message
using the [>] key and repeat the operation from step 2.
(2) Selection of Contact Output Operation (open/closed) (FUNCTION
"C" to FUNCTION "F")
Select open or closed for the contact output when operated. Table 6.13 shows the types of
contact outputs and corresponding functions.
Table 6.13 Contact Outputs and Operating Functions
Contact output
FUNCTION Output terminal Factory setting
Maintenance contact output
C
M1, M2
Closed
FAIL contact output
D
F1, F2
Open
High and low alarm contact output
E
C1, C2
Open
Contact output during automatic
cleaning and automatic zero calibration
F
C1, C2
Closed
The following shows an example of the setting operations for the maintenance contact output.
Similar operations apply to the other settings.
(Example of display)
1. Select FUNCTION "C" in the <PROGRAM2>
=> _ _1 (Open)
mode. The current set status is displayed.
2. Press the [>] key to select the setting codes.
=> _ _2 (Closed)
F6701.ai
_ _1
_ _2
(Open)
(Closed)
3. Press the [ENT] key to enter the setting.
=> _ _2
The
display
fl
ashes once.
(3) Selection of Execution/Stop of the Failure Detection Functions
(FUNCTION "1." to FUNCTION "9.")
These functions can select execution/stop of part of the failure detection functions. Table 6.14
shows the detected failures and their operation functions.
3rd Edition : May. 31, 2010-00