<7. MAINTENANCE>
7-2
IM 12E04A02-02E
7.3 Cleaning of Measuring Cell
Clean the inside of the detector directly with a brush or the like.
1. Select the <MAINT.> mode.
2. Select the draining status (for the statuses of each value during draining, see Table 6.15).
3. Loosen screw A under the detector cover and open the cover (see Figure 7.1).
4. After the water in the detector is drained off, wipe the inside of the detector with a soft brush,
sponge, or the like. Be careful not to mar or scratch the inside and not to splash the lenses
in the light source and light-receiving part with water.
5. Open valves V4 and SV2 to supply cleaning water and discharge the dirt with the water.
Also in this case, gradually open valve V4 so as not to splash the lenses in the light source
and light-receiving part with water.
6
Tighten screw A after cleaning is completed and execute zero and span calibrations.
7. Finished.
F0701.ai
Screw A
Measuring cell
Light source
Light-receiving part
Figure 7.1 Detector
3rd Edition : May. 31, 2010-00