<5. FUNCTIONS>
5-7
IM 12E04A02-02E
V:
Measuring signal, a voltage signal from the detector
T2:
Turbidity value after correction
K:
Sensitivity correction factor
This is calculated in the calibration of item (5) in Table 5.4, and can be displayed and changed using item (12) in
Table 5.4.
The permissible range in items (5) and (12) in Table 5.4 is 0.25 <= K <= 4.00.
B: Zero-correction
coef
fi
cient
This is calculated in item (3) in Table 5.4 and can be displayed and changed using item (11) in Table 5.4.
-9 mg/l <= B <= 9 mg/l is the permissible range for items (3) and (11) in Table 5.4.
(Note 1) Turbidity in the <MEAS.> mode is displayed in T2 (turbidity after correction). However, when the turbidimeter is shipped from
the factory, the zero and sensitivity correction factors are 0.0 mg/l and 1, respectively. Thus, T1 = T2.
(Note 2) For zero and span calibration of (1), (2), (4), and (8) referencing the standard solution, T1 (turbidity before correction) is
always displayed as turbidity.
5.4 Functions Related to Display
In the <MEAS.> mode, each item in Table 5.5 can be selectivity displayed. In addition, Table 5.6
shows the functions related to turbidity display.
Table 5.5 Functions Related to Display
Function
Description
Mode
FUNCTION
(1) Turbidity display
*1
Turbidity is displayed.
MEAS.
1
(2) Lamp voltage
display
The lamp voltage in the converter is displayed. *1
MEAS.
2
(3) Measured signal
current display
The measured value by the receiving element that
measures scatterd light is displayed in the current value.
MEAS.
3
(4) Reference signal
voltage display
The signal of the receiving element that controls tha lamp
light quantity is displayed in the voltage value.
MEAS.
4
(5) Analog output %
display
The analog output is displayed after being converted into
% FS.
MEAS.
5
(6) Analog output mA
display
The analog output is displayed after being converted into
4 to 20 mA.
MEAS.
6
(7) Zero error display
The measured signal in zero calibration does not
become 0 mV exactly. This error (mV) is displayed after
being converted to turbidity.
MEAS.
7
(8) Slope display
Sensitivity at the latest span calibration to the sensitivity
at reference sensitivity calibration is displayed in %.
MEAS.
8
(*1)
Owing to item 3 in Table 5.6, the unit indication can be changed.
Table 5.6 Functions Related to Turbidity Display
Function
Description
Mode
FUNCTION
(1) Negative turbidity
display hold *1
If the value of the turbidity is negative, the turbidity is
fi
xed at 0.00 mg/l.
PROGRAM 1
7.
(2) Display hold
automatic
cleaning or zero
calibration
The display in the automatic cleaning or zero calibration
is
fi
xed.
PROGRAM 1
8.
(3) Selection of
turbidity unit
indication
The unit of turbidity indication (mg/l, degree, and FTU)
can be selected.
PROGRAM 1
C.
(*1)
Even if this function is on, a negative value is displayed in other than the <MEAS.> mode.
5.5 Functions Related to Contact Output
An on or off operation can be selected for every contact output.
The output AUX (C1, C2) is used for either upper and lower limit alarm output or automatic
cleaning and calibration output.
3rd Edition : May. 31, 2010-00