<6. PROCEDURES FOR CONVERTER OPERATION>
6-24
IM 12E04A02-02E
(8) Setting Turbidity Value of Calibration Plate (FUNCTION "A")
Upon shipment from the factory, the value of the attached calibration plate has been set.
The following shows the setting key operations.
(Example of display)
1. Select FUNCTION "A" in the <PROGRAM1>
=> 0090.0 mg/l
mode.
2. Press the [>] key.
=> 0090.0 mg/l
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set it to the
=> 0095.0 mg/l
desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the set value.
=> 0095.0 mg/l
The entire display
fl
ashes once.
In step 4, if the set value is beyond the permissible range (0 to 2000 mg/l), "not" is displayed.
Reset the message by pressing the [>] key and repeat the operations from step 2.
(9) Setting Zero-Correction Factor (FUNCTION "B")
The zero-correction factor obtained in zero-shift calibration (zero correction) (FUNCTION "3" in
<MAINT.> mode) can be displayed or directly changed. For details on the zero-shift calibration
function, see Section 5.3 (*3).
The following shows the setting key operations.
(Example of display)
1. Select FUNCTION "B" in the <PROGRAM1>
=> 00.000 mg/l
mode.
If the display is only to be con
fi
rmed,
operations of 2, 3, and 4 are not necessary.
2. Press the [>] key.
=> 00.000 mg/l
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set it to the
=> 00.050 mg/l
desired value using the [^] or [v] key.
The
fl
ashing digit moves to theright
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the set value.
=> 00.050 mg/l
The entire display
fl
ashes once.
In step 4, if the set value is beyond the permissible range (±9 mg/l), "not" is displayed.
Reset the message by pressing the [>] key and repeat the operations from step 2.
(10) Setting Sensitivity Correction Factor (FUNCTION "C")
The sensitivity correction factor obtained in span calibration with sensitivity correction
(FUNCTION "5" in <MAINT.> mode) can be displayed or directly changed. For details on the
span calibration function, see Section 5.3 (*3).
3rd Edition : May. 31, 2010-00