<2. SPECIFICATIONS>
2-5
IM 12E04A02-02E
2.3 Model and Suf
fi
x Codes
(Style code: S3)
Model
Suf
fi
x code
Option code
Description
TB400G
······················· ··················· Open-surface turbidimeter using scattered light
Output
–4
–5
···················
···················
4 to 20mA DC
1 to 5V DC
Power supply
–1
–3
–6
–7
···················
···················
···················
···················
100 V AC, 50/60 Hz
110 V AC, 50/60 Hz
200 V AC, 50/60 Hz
220 V AC, 50/60 Hz
Samplyng system,
automatic cleaning, and
automatic zero calibration
–NN
–A1
–A2
–A3
···················
···················
···················
···················
Without sampling system, automatic cleaning, or automatic zero
calibration *1
With sampling system, but without automatic cleaning or
automatic zero calibration
With sampling system and automatic cleaning, but without
automatic zero calibration
With sampling system, automatic cleaning, and automatic zero
calibration
Optional speci
fi
cation
/P
/R
/B
/S
/L
/SCT
/PHN
/PHU
/FC
/ARS
/PSL
Pipe mounting bracket *2
Rack mounting bracket *2
Rear piping installation *3
Stainless steel stanchion *3
Spike detection function for low range *3 *4
Stainless steel tag plate
Build-in pH meter without ultrasonic cleaning function *3 *5
Build-in pH meter with ultrasonic cleaning function *3 *5
Built-in non-reagent type free available chlorine analyzer *3 *5
With arrester *6
Calibration using polystyrene latex *7
*1
Provide a defoaming tank (head tank) so that the
fl
ow rate of the water sample reaches the speci
fi
ed
fl
ow (1.5 to 2 l/min).
*2 Optional
speci
fi
cations for selecting “without sampling system (code: -NN)”
*3 Optional
speci
fi
cations for selecting “with sampling system (code: -A1, -A2, or -A3)”
*4
For applications where the measuring range is low (200 mg/l or less) and air bubbles are likely to occur, add the “L” option (spike
detection function). This option is not available for higher measuring range.
*5
A pH meter or non-reagent type free available chlorine analyzer should be purchased separately when the relevant option code
is speci
fi
ed. You can select only one optional speci
fi
cation in the /PHN or / PHU option or the /FC option on the one turbidimeter.
*6
Option code /ARS cannot be selected when /PHN or /PHU or /FC option is speci
fi
ed.
*7
Polystyrene latex of 2 degrees is used as standard. Polystyrene latex with a size of up to 100 degrees may be handled as
Tokuchu. Kaolin is used as turbidity standard for calibration unless the /PSL option is speci
fi
ed.
2.4 Accessories
Item
Quantity
Remarks
Standard calibration plate
1
Housed in the converter
Silicone cloth
1
Lamp
1
As a spare
Fuse
4 each
1A and 3A (as spares)
Pipe mounting bracket (optional)
1 set
When specifying option code "/P"
Rack mounting bracket (optional)
1 set
When specifying option code "/R"
Soft PVC tube, 1 m
2
For detector piping (without sampling system)
Clamp
2
For detector piping (without sampling system)
3rd Edition : May. 31, 2010-00
Содержание Vigilant Plant EXA TB Series
Страница 49: ...Blank Page ...
Страница 59: ...Blank Page ...
Страница 119: ...Blank Page ...
Страница 125: ...Blank Page ...
Страница 127: ...Blank Page ...
Страница 133: ...Blank Page ...